Processes calibrated one-dimensional complex NMR free-induction decays with nmrglue into phased spectra, peak candidates, and signed integration regions. Use for raw 1D NMR processing, ppm-axis verification, apodization, Fourier transformation, manual phasing, baseline correction, or reproducible spectral integration.
cp -r skills/nmrglue ~/.muse/skills/nmrglue # or your agent's skill dir
Works with Claude Code, Codex, Gemini CLI and any agent that loads SKILL.md.Use for a uniformly sampled complex 1D FID whose acquisition parameters and complex frequency convention are available. The helper produces a descending ppm spectrum, positive peak candidates, signed region integrals, and a reproducible processing report. It does not identify compounds or assign resonances.
The executable accepts a NumPy .npz containing exactly one complex fid array or a
canonical 1D complex time-domain NMRPipe file. NMRPipe reading is tested with a synthetic
write/read round trip, known-spectrum recovery, and a small upstream NMRPipe-generated
binary fixture. Experimental Bruker, Varian, and JEOL
imports are not verified by this suite. For those formats, first inspect the relevant
nmrglue reader and acquisition metadata. Opening a converted file does not validate the
original acquisition decoding.
Read references/acquisition-and-validation.md
for conversion boundaries, axis calibration, and quantitative limits.
-i
means a resonance at offset f = (ppm - carrier_ppm) * observation_mhz has time
dependence exp(-2*pi*i*f*t). Select +i only for the opposite convention; the helper
conjugates it before processing. Validate with a known reference peak.0.5 is suitable for the supplied causal synthetic example;
acquisition and prior preprocessing may require another value.phase0_deg + phase1_deg * index / zero_fill_points is applied after FT;
index zero is the high-ppm edge. There is no implicit pivot or automatic phase estimate.baseline to linear and add baseline_regions_ppm containing at least two
regions. Inspect residuals and broad peaks; fitting through signals biases integrals.Tested with Python 3.12, nmrglue 0.12, NumPy 2.5.3, and SciPy 1.18.1:
uv run --no-project --python 3.12 --with nmrglue==0.12 --with numpy==2.5.3 --with scipy==1.18.1 \
python skills/nmrglue/scripts/process_1d.py fid.npz processing.json nmr-result
Paths assume the collection root. Adjust them when installed elsewhere. The output directory must be new, so repeated processing keeps previous results reviewable.
For an existing 1D NMRPipe FID, add --input-format nmrpipe and supply its path in place
of fid.npz. The helper requires the canonical FDF2 direct dimension, complex quadrature,
a time-domain flag, and agreement between header and JSON spectral width, observation
frequency, and carrier. JSON settings remain explicit; a mismatch fails instead of silently
recalibrating. FDF2TDSIZE must equal the stored complex-point count, and FDF2CENTER /
FDF2ORIG must describe a canonical centered axis. Previously zero-filled, truncated,
or recentered files need a separate acquisition-aware workflow. The nucleus/complex sign
and previous digital-filter corrections still need acquisition evidence. A time-domain
flag alone does not establish an unprocessed FID.
This executable synthetic example matches the supplied settings, generates resonances at 3 and 7 ppm in a 1:2 amplitude ratio, and does not represent an experimental sample:
import numpy as np
t = np.arange(8192) / 4000.0
fid = sum(a * np.exp(-np.pi * 2.0 * t)
* np.exp(-2j * np.pi * (ppm - 5.0) * 400.0 * t)
for ppm, a in [(3.0, 1.0), (7.0, 2.0)])
np.savez("fid.npz", fid=fid)
Run it with assets/processing.json as the settings argument. The repository suite
executes this signal and the CLI, checks both peak locations within 0.001 ppm, checks
integral ratio and analytic area, and checks phase and baseline recovery. The NMRPipe
round-trip test writes this FID using ng.pipe.create_dic/ng.pipe.write, reads it through
the CLI, and verifies the recovered peaks and integral ratio. Processed frequency-domain
files and conflicting calibration metadata are rejected.
The 2,176-byte upstream fixture checks complex sample order and header calibration using
a file generated by NMRPipe's simTimeND / SET tools. Those native tools were not run
in this review; this is fixture compatibility, not a live NMRPipe processing comparison.
spectrum.csv: descending ppm, real signal after baseline correction, phased imaginary
signal, and the fitted real baseline. Plot NMR with the high-ppm end on the left.report.json: input/settings SHA-256, package versions, all settings, acquired duration,
zero-filled digital spacing, positive peak candidates, and signed region areas.Integrals use endpoint interpolation and trapezoidal integration along increasing ppm; area units are arbitrary signal times ppm, independent of display direction. Regions outside the sampled ppm axis fail rather than being silently clipped. Peak prominence is a fraction of the largest positive real intensity; it is not a noise-derived detection limit. Strong solvent signals can obscure weak candidates at the default threshold.
For quantitative NMR, additionally establish relaxation delay, pulse angle, saturation, receiver behavior, internal/external reference amount, and integration uncertainty. The helper does not calculate concentrations or correct unequal relaxation. Preserve these limits with the result rather than converting arbitrary areas to molecule counts.
The hosted latest documentation identified itself as 0.9-dev when checked; the actual
0.12 package APIs and numerical behavior were tested. Its proc_base.fft uses the
negative-exponent NumPy FFT followed by fftshift; NMRPipe's FT convention corresponds
to fft_positive, so do not substitute it without revisiting the FID sign and phase.
See the v0.12 processing source.
Multidimensional processing, nonuniform sampling, automated assignment, and experimental vendor imports remain outside
this helper's validated scope.